Logo of The Journal of Bone & Joint Surgery (Br)
Quick search:        
          Advanced Search
Guest Access | Sign In
Journal of Bone and Joint Surgery - British Volume, Vol 85-B, Issue SUPP_II, 101.  
Copyright © 2003 by British Editorial Society of Bone and Joint Surgery
This Article
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Solan, M C
Right arrow Articles by Jinnah, R H
Right arrow Search for Related Content
PubMed
Right arrow Articles by Solan, M C
Right arrow Articles by Jinnah, R H

British Orthopaedic Association


Birmingham – 12–14 September, 2001

President – Professor Charles Galasko


THE MAYO CONSERVATIVE FEMORAL STEM: A PHOTO-ELASTIC COATING STUDY TO EVALUATE STRAIN IN THE PROXIMAL FEMUR

M C Solan; B Parks; and R H Jinnah

7 Monks Walk, Reigate, Surrey RH2 OSS

The Mayo Conservative uncemented stem (Zimmer, Warsaw, USA) is designed to conserve proximal bone stock by virtue of a minimal neck resection and to maintain proximal femoral stress transfer, thereby reducing problems associated with stress shielding.

This study was performed to evaluate proximal femoral strain after implantation of the Mayo stem, in cadaveric femora.

Eight fresh-frozen cadaveric femora (each selected at random from within a pair) of known bone mineral density were prepared and coated with photoelastic materials (Measurements Group, Raleigh NC). Strain patterns of the intact bone were determined using a reflection polariscope, and recorded photographically, while under load. Quantitative measurements were taken from set points of the proximal femur. The femoral head was then replaced using a Mayo femoral prosthesis. Under the same loading conditions strain patterns were re-examined and measurements taken from the same set points.

The strain patterns following insertion of the Mayo stem closely matched those seen in intact femora except in two areas. Strain was reduced in the region of the lesser trochanter (53% of normal), although more proximal than this strain in the neck was closer to intact values (78% of normal). Previous studies have found that implantation of diaphyseal press fit stems in particular have led to significant reductions in shear strain values in the calcar region and distally along the medial border of the femur.

This study documents the strain pattern in the proximal femur after implantation with a new "conservative" short stem cementless prosthesis. The hypothesis that the Mayo femoral stem maintains proximal femoral stress transfer and may thus prevent stress shielding in vivo remains to be proven, but is supported by the results of this study.

The abstracts were prepared by Mr Simon Donell. Correspondence should be addressed to him at the Department of Orthopaedics, Norfolk & Norwich Hospital, Level 4, Centre Block, Colney Lane, Norwich NR4 7UY, United Kingdom






(c) British Editorial Society of Bone and Joint Surgery All Rights Reserved
Registered charity no: 209299     Print ISSN: 0301-620X
Hip, Knee, Trauma, Upper limb, Foot & Ankle, Paediatrics, Oncology, Spine, Arthroplasty, General