Logo of The Journal of Bone & Joint Surgery (Br)
Joint Replacement Instrumentation Limited (JRI) Ad
Quick search:        
          Advanced Search
Guest Access | Sign In
Journal of Bone and Joint Surgery - British Volume, Vol 88-B, Issue SUPP_II, 208.  
Copyright © 2006 by British Editorial Society of Bone and Joint Surgery
This Article
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Böhling, U.
Right arrow Search for Related Content
PubMed
Right arrow Articles by Böhling, U.

European Rheumatoid Arthritis Surgical Society


Pfäffikon, Switzerland – 25–27 May, 2006

President – Beat R. Simmen


S18 BIONIC SYSTEM IN TOTAL HIP ARTHROPLASTY

BöhlingU.

Berlin

Summary: The tribological development in the metal-metal partners brought out a new surface texture for the femoral cup. A reduction of the coefficient of friction and a reduction of the metallic abrasion on almost half are secured in lab tests. First clinical applications took place in our hospital.

Question: Total hip surface replacement in metal-metal partners have, although in small quantity, metallic abrasion, which leads to clear increases of the serum concentration of the serological level. The long-term effect on the organism is not so far clarified. A reduction of the abrasion quantity is however worthwhile. Is a structural change of the surface of the femoral cup suitable to cause such a reduction of the metallic abrasion?

Methodology: Derived from bionic systems in nature the surface of total hip surface replacement was changed by dimple like surface, so that in presence of liquid an accumulation of liquid between both sliding partners takes place and takes place thus on this liquid film the articulation. Laboratory tests on a simulator were accomplished, in order to examine whether the desired reduction of the abrasion is obtained. This surface is called Biosurf-surface.

Results: Simulator wear tests of the standard total hip surface replacement in the metal-metal partner have been compared with the Biosurf-surface. The simulator attempts showed that after 5 million motion cycles the Biosurf –system the abrasion quantity in milligram reduces to more than a half. By this clear reduction of particles the concentration of the metal ions which can be expected in the serum is clearly smaller, so that thereby a contribution is made to reduce the possibly damaging influence of metallic ions in the organism.

Correspondence should be addressed to ERASS Office, Schulthess Klinik, Lengghalde 2, CH-8008 ZURICH, Switzerland.