Logo of The Journal of Bone & Joint Surgery (Br)
Joint Replacement Instrumentation Limited (JRI) Ad
Quick search:        
          Advanced Search
Guest Access | Sign In
Journal of Bone and Joint Surgery - British Volume, Vol 88-B, Issue SUPP_II, 212.  
Copyright © 2006 by British Editorial Society of Bone and Joint Surgery
This Article
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Hilker, A.
Right arrow Search for Related Content
PubMed
Right arrow Articles by Hilker, A.

European Rheumatoid Arthritis Surgical Society


Pfäffikon, Switzerland – 25–27 May, 2006

President – Beat R. Simmen


S37 PROBLEMATIC EARLY RESULT WITH THE PYROCARBON MCP-ENDOPROSTHESIS ACC. TO BECKENBAUGH

HilkerA.

Wetter

Background The AscensionTM MCP-implant acc. to Beckenbaugh is an unconstrained endoprosthesis consisting of a graphite core with a pyrocarbon coating. Long term results published by the Mayo Clinic gave rise to the hope of good ingrowth properties and a long durability due to the optimum wear resistance of the material. Purpose of this study was the evaluation of the early results of this implant applied on the patients of a rheumatoid arthritis surgery department.

Methods In a prospective manner we evaluated the data of 18 patients with 19 involved hands in which we implanted a total of 29 Ascension endoprostheses between July, 2000 and Oktober, 2002. The mean follow up time was 39.7 months. We assessed hand function (SODA-score), mobility, radiographic appearance and the subjective parameters pain, functionality, power, cosmetic appearance and global satisfaction.

Results The SODA score increased from 82.8 preoperatively to 92.9 (maximum 108) at follow up. With 50.9 and 51.3 deg. respectively the range of motion remained almost unchanged. On a visual analogue scale from 0 – 10 the subjective results were: pain 2.1 (–5.2), functionality 6.5 (+3.0), power 5.3 (+2.3), cosmetic appearance 8.0 (+3.4), global satisfaction 8.1.

Radiographically we observed suspicious radiolucent lines around 17 of 58 implanted components (29%), three of them progressive. Around the tips of three stems an osteolytic area with minimal progression was found. Three stems showed a slight migration. The following intraoperative complications were observed: Two bone fissures and three lacerations of a collateral ligament, which all healed uneventfully. Five endoprostheses were converted into a flexible hinge silicone implant for persistent pain and loosening between 12 and 48 months p.o. One more implant had to be revised due to a late haematogenous infection 4 years p.o.

Conclusion With the AscensionTM MCP implant a good pain relief can be obtained in almost every case but only rarely an improvement in the range of motion can be seen. The subjective results were predominantly good. But the apparently inconstant bony ingrowth resulted in a rate of revisions for aseptic loosening as high as 17% at a mean follow up of 3 1/2 years. A modification of the implant leading to a better bony fixation would be required before a wider clinical use of this implant can be recommended.