Logo of The Journal of Bone & Joint Surgery (Br)
Quick search:        
          Advanced Search
Guest Access | Sign In
Journal of Bone and Joint Surgery - British Volume, Vol 90-B, Issue SUPP_I, 7-8.  
Copyright © 2008 by British Editorial Society of Bone and Joint Surgery
This Article
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Gupta, S.
Right arrow Articles by Chatterji, S
Right arrow Search for Related Content
PubMed
Right arrow Articles by Gupta, S.
Right arrow Articles by Chatterji, S

British Orthopaedic Association


Birmingham, England: 17–19 September 2003

President: Professor Paul Gregg


ACCORD (JOHNSON-ELLOY) TOTAL KNEE REPLACEMENT: WHY DID IT FAIL?

SK Gupta; KJJ Tayton; C Dent; and S Chatterji

West Wales Hospital, Carmarthen SA31 2AF

To study the survival analysis of the Accord TKR and to analise the causes for its failure. 111 Accord knees were implanted in 106 patients between 1986 and 1996. All components were cemented. Eighty-seven patients were followed up and assessed according to the Knee Society Clinical Rating System.

Life table analysis of this implant using revision as the end point shows a survival of 25% at 11–12 years.

Of the 87 knees, only 31 were still in situ and of these 7 showed radiographic signs of severe loosening. However, the average knee score for these 31 was only 65/100 and average functional score was down to 42/100.

56 implants have been revised, 21 due to aseptic loosening, 11 due to gross valgus/varus instability, 9 due to deep infection, 8 due to loosening of the patella liner and 7 due to mobile bearing complications.

All 21 (24%) cases of aseptic loosening were found to have a loose femoral component; however, 56% of the total showed significant radiographic osteolysis around the stem of the tibial implants.

Retrieved implants in 5 patients showed significant delamination of the UHMWP at its margins and also revealed a track through the tibial baseplate into the medullary cavity of the tibia.

Conclusion: Early failure of this implant is due to increased debris formation from the UHMWP due to edge loading and early delamination probably as a consequence of the shape of its articulating surface. This situation was significantly aggravated by a design fault in the tibial baseplate, which encouraged unrestricted access of debris into the medullary cavity of the tibia and hence early and severe osteolysis.